Hi-2621 pH and Orp Benchtop Meter with Data Logging

The HI-2621 benchtop meter is a fast, accurate pH and ORP meter with probes for pH and Orp.   The sensors measure the parameter together with the temperature.


Simply and easy to use : Configuration, calibration, measurement and data logging.


Useful for routine applications together with a Standard Operating Mode, enabling all  features and capabilities.


The HI-2621 meter has a large display - 5.5” (14 cm) LCD.


Main feature points

  • Automatic parameter recognition
  • pH/ORP choice of measurement units: pH, mV;
  • CAL Check indicators
  • Sensor Check indicators
  • Basic mode for simplified operation
  • Dedicated GLP key
  • GLP data included with logged data
  • Simplified data transfer to a PC
  • Internal clock and date

Key Features

  • Digital probe input
  • Two operating modes - Standard or Basic Operating Modes (pH and EC only). Standard Mode enables all features while Basic Mode displays a simplified screen and features
  • Data storage up to 1000 log records.
  • GLP - The sensor retains data from the most recent calibration, including date, time, and buffers utilised.
  • CAL Check -  CAL Check feature analyses the pH electrode’s response in calibration buffers to detect potential issues such as contaminated solutions or dirty electrodes.
  • ORP measurement - Measure ORP (REDOX) in mV units by using a compatible ORP sensor.
 pH orp data logging benchtop meter
Specifications of the HIO2621
pH - using HI-11310 Electrode pH
_____
pH Range basic mode: -2.00 to 16.00 pH; standard mode: -2.000 to 16.000 pH
Resolution basic mode: 0.01 pH; standard mode: 0.001 pH
Accuracy basic mode: ±0.01 pH; standard mode: ±0.002 pH
Calibration Basic Mode: Automatic, up to three points, option to select 5 standard buffers
Standard mode: Automatic, up to five point calibration, 7 standard buffers available and two custom buffers.
Temperature Compensation Automatic, -5.0 to 100.0°C (using integral temperature sensor)
Logging Up to 1000 (in Standard Mode) records organised in:
Manual log-on-demand, maximum 200 logs
Manual log-on-stability, maximum 200 logs
Interval logging in Standard Mode
________
mV - using HI-11310 pH electrode
_____
mV pH Range ±1000 mV
Resolution 0.1 mV
Accuracy at 25°C ±0.2 mV
__________
ORP – using HI-36180 ORP electrode
_____
Orp Range ±2000.0 mV
Resolution 0.1 mV
Accuracy at 25°C ±0.2 mV (±999.9mV); ±1mV (±2000mV)
Calibration Single point offset
Temperature Compensation Automatic, using integral temperature sensor -5.0 to 100.0°C
Logging Up to 1000 in Standard Mode records organised in:
Manual log-on-demand, maximum 200 logs
Manual log-on-stability, maximum 200 logs
Interval logging in Standard Mode
Temperature Range -20.0 to 120.0 °C (temp limits reduced to probe limits)
Resolution 0.1°C
Accuracy ±0.5°C